Paper
8 February 1985 Intercomparison Between Independent Irradiance Scales Based On Silicon Photodiode Physics, Gold Point Blackbody Radiation, And Synchrotron Radiation
A. R. Schaefer, R. D. Saunders, L. R. Hughey
Author Affiliations +
Abstract
A radiometer composed of a silicon photodiode, interference filter, and integrating sphere was characterized and calibrated against an absolute silicon detector standard at 600 nm using a CW dye laser. This radiometer was then used to measure the irradiance at 600 nm from spectral irradiance lamps calibrated against a gold point blackbody, and also the irradiance from the NBS electron storage ring, SURF II. These results were intercompared with the independently calculated irradiances from these two sources, with overall agreement of better than one percent. Various aspects of the measurements are discussed.
© (1985) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
A. R. Schaefer, R. D. Saunders, and L. R. Hughey "Intercomparison Between Independent Irradiance Scales Based On Silicon Photodiode Physics, Gold Point Blackbody Radiation, And Synchrotron Radiation", Proc. SPIE 0499, Optical Radiation Measurements, (8 February 1985); https://doi.org/10.1117/12.971069