5 June 2003 Evaluation of reconstruction accuracy of cone beam volume CT breast imaging for different scanning orbits
Ruola Ning, Xianghua Lu, Biao Chen, Yong Yu
Author Affiliations +
Abstract
Conventional screen film mammography is the most effective tool for the early detection of breast cancer currently available. However, conventional mammography has relatively low sensitivity for detecting small breast cancers (under several millimeters). Specificity and the positive predictive value of mammography remain limited owing to an overlap in the appearance of benign and malignant lesions, and surrounding structure. We propose to address the limitations accompanying conventional mammography by incorporating a cone beam volume CT (CBVCT) reconstruction technique with a recently developed flat panel detector. In this study, we present the results obtained from a computer simulation study and a breast-imaging phantom experimental study to find out the reconstruction accuracy of different cone beam volume scanning orbits for CBVCTBI, and to determine if different partial scan protocols (less than 360°) are appropriate for breast cancer detection. Three types of CBVCTBI scanning orbits were simulated using an uncompressed breast phantom. The reconstruction accuracy was evaluated as a function of different scanning orbits assuming reconstruction with 360° was the gold standard. The results indicate that with 180° plus cone angle orbit, the reconstruction error is below 4% that is in the acceptable range. In addition, a preliminary phantom study using both 360° and 180° plus cone angle orbits, was conducted on the current flat panel detector-based CBVCT prototype scanner. The companion CBVCT reconstruction images of an uncompressed breast-image phantom are presented.
© (2003) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Ruola Ning, Xianghua Lu, Biao Chen, and Yong Yu "Evaluation of reconstruction accuracy of cone beam volume CT breast imaging for different scanning orbits", Proc. SPIE 5030, Medical Imaging 2003: Physics of Medical Imaging, (5 June 2003); https://doi.org/10.1117/12.480252
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Breast

Computer simulations

Mammography

Sensors

Breast cancer

Data acquisition

Breast imaging

RELATED CONTENT


Back to Top