Paper
15 May 2003 Windmill artifact in multislice helical CT
Author Affiliations +
Abstract
Multi-slice helical CT-systems suffer from windmill artifacts: black/white patterns that spin off of features with high longitudinal gradients. The number of black/white pairs matches the number of slices (detector rows) in the multi-slive detector. The period of spin is the same as the helical pitch. We investigate the cause of the pattern by following the traces of selected voxels through the multi-slive detector array as a function of view position. This forms an "extracted sinogram" which represents the data used to reconstruct the specific voxel. Now we can determine the cause of the artifact by correlating the windmill streak in the image with the extracted data. The investigation shows that inadequate sampling along the longitudinal direction causes the artifact.
© (2003) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Michael D. Silver, Katsuyuki Taguchi, Ilmar A. Hein, BeShan Chiang, Masahiro Kazama, and Issei Mori "Windmill artifact in multislice helical CT", Proc. SPIE 5032, Medical Imaging 2003: Image Processing, (15 May 2003); https://doi.org/10.1117/12.483585