Paper
20 November 2003 Wavelength stability of sweepmeter-based tunable laser spectrometer
Author Affiliations +
Proceedings Volume 5278, Sixth Pacific Northwest Fiber Optic Sensor Workshop; (2003) https://doi.org/10.1117/12.544458
Event: Sixth Pacific Northwest Fiber Optic Sensor Workshop, 2003, Troutdale, OR, United States
Abstract
Optical fiber sensors must compete in performance with traditional electronic sensors, such as quartz crystal pressure and temperature monitors. The precision of commercial electronic sensors can reach the parts-per-billion (ppb) level. To test the precision of a laser based spectrometer system, repeated measurements of an absorption line of a molecular gas cell were made. The Allan deviation is computed, and it is shown that the laser interrogation system, built completely out of commercially available components, can achieve precision at the 10-ppb level.
© (2003) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Kurt R. Vogel and Christopher John Myatt "Wavelength stability of sweepmeter-based tunable laser spectrometer", Proc. SPIE 5278, Sixth Pacific Northwest Fiber Optic Sensor Workshop, (20 November 2003); https://doi.org/10.1117/12.544458