30 September 2004 Dynamic alignment of a Michelson interferometer using a position-sensitive device
Carole G. Prevost, Jerome Genest
Author Affiliations +
Abstract
The objective of this project is to dynamically align a Michelson interferometer. This interferometer is designed to measure the spatial coherence of light sources. In order to obtain accurate coherence measurements, many parameters within the interferometer must be controlled with high precision. Once such parameter is the angle of inclination between mirrors in the interferometer. To introduce a well controlled and variable amount of tilt between the mirrors, we propose here a new dynamic alignment method employing a position sensitive device. A theoretical angular precision of 1 microrad was desired and an experimental angular precision of 2 microrads was obtained.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Carole G. Prevost and Jerome Genest "Dynamic alignment of a Michelson interferometer using a position-sensitive device", Proc. SPIE 5528, Space Systems Engineering and Optical Alignment Mechanisms, (30 September 2004); https://doi.org/10.1117/12.562611
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.