26 October 2004 Phase tomography using an x-ray Talbot interferometer
Atsushi Momose, Shinya Kawamoto, Ichiro Koyama, Yoshio Suzuki
Author Affiliations +
Abstract
X-ray phase tomography with X-ray Talbot interferometry (XTI) is reported. XTI employs two transmission gratings and generates a contrast corresponding to the differential phase shift caused by a sample. Quantitative phase measurement and tomographic image reconstruction with XTI are demonstrated for biological samples. Finally, the possibility of medical applications of XTI is discussed, based on the advantage of XTI that divergent and polychromatic X-rays are available.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Atsushi Momose, Shinya Kawamoto, Ichiro Koyama, and Yoshio Suzuki "Phase tomography using an x-ray Talbot interferometer", Proc. SPIE 5535, Developments in X-Ray Tomography IV, (26 October 2004); https://doi.org/10.1117/12.557572
Lens.org Logo
CITATIONS
Cited by 35 scholarly publications and 20 patents.