You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
19 September 1985Signal-To-Noise-Ratio Measurements On Two High-Resolution Screen-Film Systems
Measurements of the performance of two high-resolution radiographic screen-film systems at 50 and 30 kVp are given in terms of the noise equivalent number of recorded x-ray quanta (NEQ) and the detective quantum efficiency (DQE). Additionally, the MTF, contrast transfer function (CTF), and noise power spectra (NPS) of these imaging systems are presented. The NEQ, DQE, CTF, and NPS are shown to be strong functions of both spatial frequency and exposure. The shapes of these surfaces are significantly different from those published for general-purpose radiographic systems; these data provide a useful context within which screen-film system performance can be interpreted.
The alert did not successfully save. Please try again later.
Phillip C. Bunch, Kenneth E. Huff, "Signal-To-Noise-Ratio Measurements On Two High-Resolution Screen-Film Systems," Proc. SPIE 0555, Medical Imaging and Instrumentation '85, (19 September 1985); https://doi.org/10.1117/12.949478