Paper
20 January 2005 Influence of pinhole size and fiber optic imaging bundle on resolution in confocal endoscopic imaging system
Huayong Ge, Qiushi Ren, Baohua Wang, Wangrong Li
Author Affiliations +
Abstract
A novel confocal endoscopic imaging system is designed for imaging thick sample. The lateral and axial detected image intensity is deduced and calculated based on point spread function (PSF). The analysis of influence of pinhole and fiber on resolution is given. Our researches show that the lateral resolution improvement is much more sensitive to pinhole size than the axial resolution. Confocal system has narrower half width at half maximum (HWHM) and higher lateral and axial resolution for smaller pinhole. At last a criterion for pinhole size is introduced first of all to ensure confocal imaging and maximize signal-to-noise ratio, which is very helpful in the practical design of confocal endoscopic imaging system.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Huayong Ge, Qiushi Ren, Baohua Wang, and Wangrong Li "Influence of pinhole size and fiber optic imaging bundle on resolution in confocal endoscopic imaging system", Proc. SPIE 5633, Advanced Materials and Devices for Sensing and Imaging II, (20 January 2005); https://doi.org/10.1117/12.575698
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Confocal microscopy

Imaging systems

Image resolution

Endoscopy

Fiber optics

Microscopes

Signal to noise ratio

Back to Top