Paper
13 April 2005 Curve fitting for quantitative measurement of attenuation coefficients from OCT images
Author Affiliations +
Abstract
We fit different analytical models to the OCT signal obtained from scattering Intralipid solutions in order to extract the scattering coefficient. These models are based on both single scattering models and multiple scattering models of the OCT signal. Using the statistical outcome of the fitting procedure we establish the most suitable model for each of the scattering coefficients μs = 2.5-26 mm-1. Our results indicate that for our specific class of OCT systems and samples/tissues, it is justified to use the single scattering model to extract the scattering coefficient from samples with scattering coefficients up to 17 mm-1.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Dirk J. Faber, Maurice C.G. Aalders, and Ton G. van Leeuwen "Curve fitting for quantitative measurement of attenuation coefficients from OCT images", Proc. SPIE 5690, Coherence Domain Optical Methods and Optical Coherence Tomography in Biomedicine IX, (13 April 2005); https://doi.org/10.1117/12.592371