You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
12 April 2005Distortion correction, calibration, and registration: toward an integrated MR and x-ray interventional suite
We present our co-registration results of two complementary imaging modalities, MRI and X-ray angiography (XA), using dual modality fiducial markers. Validation experiments were conducted using a vascular phantom with eight fiducial markers around its periphery. Gradient-distortion-corrected 3D MRI was used to image the phantom and determine the 3D locations of the markers. XA imaging was performed at various C-arm orientations. These images were corrected for geometric distortion, and projection parameters were optimized using a calibration phantom. Closed-form 3D-to-3D rigid-body registration was performed between the MR markers and a 3D reconstruction of the markers from multiple XA images. 3D-to-2D registration was performed using a single XA image by projecting the MR markers onto the XA image and iteratively minimizing the 2D errors between the projected markers and their observed locations in the image. The RMS registration error was 0.77 mm for the 3D-to-3D registration, and 1.53 pixels for the 3D-to-2D registration. We also showed that registration can be performed at a large IS where many markers are visible, then the image can be zoomed in maintaining the registration. This requires calibration of imperfections in the zoom operation of the image intensifier. When we applied the registration used for an IS of 330 mm to an image acquired with an IS of 130 mm, the error was 42.16 pixels before zoom correction and 3.37 pixels after. This method offers the possibility of new therapies where the soft-tissue contrast of MRI and the high-resolution imaging of XA are both needed.
The alert did not successfully save. Please try again later.
Luis F. Gutierrez, Guy Shechter, Robert J. Lederman, Elliot R. McVeigh, Cengizhan Ozturk, "Distortion correction, calibration, and registration: toward an integrated MR and x-ray interventional suite," Proc. SPIE 5744, Medical Imaging 2005: Visualization, Image-Guided Procedures, and Display, (12 April 2005); https://doi.org/10.1117/12.597823