Paper
3 June 2005 Dual-laser CCD-Raman spectrometer applied to on-line measurement of paper coating paste
Pentti Niemela, Juha Sumen, Janne Suhonen
Author Affiliations +
Abstract
A dual-laser Raman spectrometer was developed, in which the two lasers image two different parts of the spectrum alternately on the same spectrograph. This solves the resolution/bandwidth trade-off problem of CCD Raman spectrometers, previously solved with a dual-grating technique. The dual-laser technique has, however, an additional advantage of alleviating the contradiction between sample fluorescence and detector response, which is illustrated here by applying the technique to the measurement of paper coating paste.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Pentti Niemela, Juha Sumen, and Janne Suhonen "Dual-laser CCD-Raman spectrometer applied to on-line measurement of paper coating paste", Proc. SPIE 5826, Opto-Ireland 2005: Optical Sensing and Spectroscopy, (3 June 2005); https://doi.org/10.1117/12.623477
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Spectroscopy

Raman spectroscopy

CCD cameras

Coating

Latex

Luminescence

Carbonates

Back to Top