Germanium detectors are extensively used in astronomical instruments for far infrared observations. To meet the science objectives of future space projects, large-format far IR detectors are needed. As a first step toward this goal, we have fabricated a 2x16 Ge:Sb array with the 1x32, SB-190 CTIA cryogenic readout. The detector design as well as the preliminary results of our parametric tests are presented here. The array exhibits very good noise performance with an NEP as low as 4.0E-18 W/√Hz, and confirms the viability of the design for large format arrays.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
David Sisson, Jam Farhoomand, and Jeff Beeman "Preliminary test results of a low-noise 2x16 Ge:Sb array with a CTIA readout", Proc. SPIE 5883, Infrared Spaceborne Remote Sensing 2005, 58830O (13 September 2005); https://doi.org/10.1117/12.619003
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Detector arrays

Capacitors

Far infrared

Multiplexers

Infrared detectors

Infrared sensors

RELATED CONTENT

512 and 1024 element linear InGaAs detector arrays for...
Proceedings of SPIE (December 10 1992)
Characterization of VISTA IR detectors
Proceedings of SPIE (September 29 2004)
A 32x32 CTIA readout design for deep cryogenic applications
Proceedings of SPIE (September 07 2006)
Low-power multiplexed lead salt arrays
Proceedings of SPIE (January 05 1993)
Prototype 1x32 Ge Ga Ge Sb detector array with SBRC...
Proceedings of SPIE (December 18 2002)

Back to Top