You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
13 September 2005Preliminary test results of a low-noise 2x16 Ge:Sb array with a CTIA readout
Germanium detectors are extensively used in astronomical instruments for far infrared observations. To meet the science objectives of future space projects, large-format far IR detectors are needed. As a first step toward this goal, we have fabricated a 2x16 Ge:Sb array with the 1x32, SB-190 CTIA cryogenic readout. The detector design as well as the preliminary results of our parametric tests are presented here. The array exhibits very good noise performance with an NEP as low as 4.0E-18 W/√Hz, and confirms the viability of the design for large format arrays.