You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
28 February 2006Measurement of the spatial resolution of a clinical volumetric computed tomography scanner using a sphere phantom
The rapid development of modern multi-slice computed tomography (MSCT) scanners has provided imaging systems with cone-beam geometry, sub-millimetre slice thickness, and gantry rotation speeds approaching 0.3 seconds per revolution. Clinical MSCT scanners routinely generate volume data sets yet the methods used to quantify spatial resolution remain relatively unchanged from those used to evaluate single slice scanners. In this paper, we describe a method for quantifying the spatial resolution of an MSCT scanner, with cone-beam, geometry using a sphere phantom. By scanning a Teflon sphere embedded in a uniform silicone cylinder, the plane spread function (PlSF) and modulation transfer function (MTF) may be determined. Furthermore, the spatial resolution in the axial and trans-axial directions may be independently quantified, as well as the effects of Azimuthal blur and spiral scanning. To illustrate the utility of the sphere method, the spatial resolution of two commercially available MSCT scanners was measured.
Michael M. Thornton andMichael J. Flynn
"Measurement of the spatial resolution of a clinical volumetric computed tomography scanner using a sphere phantom", Proc. SPIE 6142, Medical Imaging 2006: Physics of Medical Imaging, 61421Z (28 February 2006); https://doi.org/10.1117/12.654969
The alert did not successfully save. Please try again later.
Michael M. Thornton, Michael J. Flynn, "Measurement of the spatial resolution of a clinical volumetric computed tomography scanner using a sphere phantom," Proc. SPIE 6142, Medical Imaging 2006: Physics of Medical Imaging, 61421Z (28 February 2006); https://doi.org/10.1117/12.654969