Paper
21 June 2006 Estimation of stapes piston motion from uniaxial interferometer measurements along observation directions at an angle with the piston axis is prone to substantial errors
Willem F. S. Decraemer, Shyam M. Khanna, Joris J. J. Dirckx
Author Affiliations +
Proceedings Volume 6345, Seventh International Conference on Vibration Measurements by Laser Techniques: Advances and Applications; 63450C (2006) https://doi.org/10.1117/12.693113
Event: Seventh International Conference on Vibration Measurements by Laser Techniques: Advances and Applications, 2006, Ancona, Italy
Abstract
We have shown that the vibration of the stapes in human, cat and gerbil exhibits all 3-D components of translation and rotation, but also that non-piston components do not contribute to the transmission path. In most experiments access for direct measurement of stapes motion in line with the piston axis is not available and piston motion is estimated from single component interferometric measurements done under observation directions that make angles up to 50° with the piston axis. We used a heterodyne microscope/interferometer to measure the vibration velocity of the stapes in human, cat and gerbil from different observation angles and calculated the complete set of 3-D motion components for the stapes. Using micro CT scans of the experimental ears, we could express the components in an intrinsic reference system and foretell the motion component that will be obtained when a single axis interferometric measurement is done under various angles with the piston axis. For low frequencies (f<2kHz) a cosine factor provides a good correction for the axis off-set, at higher frequencies the piston component can not correctly be estimated from a single off-axis observation. We will show that this may lead to serious misinterpretations of experimental results.
© (2006) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Willem F. S. Decraemer, Shyam M. Khanna, and Joris J. J. Dirckx "Estimation of stapes piston motion from uniaxial interferometer measurements along observation directions at an angle with the piston axis is prone to substantial errors", Proc. SPIE 6345, Seventh International Conference on Vibration Measurements by Laser Techniques: Advances and Applications, 63450C (21 June 2006); https://doi.org/10.1117/12.693113
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Motion measurement

Ear

Bone

Interferometers

Velocity measurements

Motion estimation

3D modeling

Back to Top