16 March 2007 In-plane artifacts in breast tomosynthesis quantified with a novel contrast-detail phantom
Tony Svahn, Mark Ruschin, Bengt Hemdal, Lars Nyhlén, Ingvar Andersson, Pontus Timberg, Sören Mattsson, Anders Tingberg
Author Affiliations +
Abstract
The purpose of this work was to develop a contrast-detail phantom that can be used to evaluate image quality in breast tomosynthesis (BT) and as a first step use it to evaluate in-plane artifacts with respect to object size and contrast. The phantom was constructed using a Polylite® resin as bulk material, as it has x-ray mass attenuation properties similar to polymethyl methacrylate (PMMA), a common phantom material in mammography. Six different materials - polyoxymethylene (POM), bakelite®, nylon, polycarbonate (PC), acrylonitrilebutadienestyrene (ABS) and polyethene (PE) - were selected to form the phantom details. For each of the six materials, five spherical objects were manufactured (diameters of 4, 8, 12, 16, and 20 mm) resulting in 30 objects that were embedded with their centres approximately aligned at the central plane of a 26 mm thick Polylite® block (210 mm x 300 mm). A 20 mm thick PMMA block was added to yield a phantom with attenuation properties similar to 45 mm PMMA that could simulate a so-called standard breast (50 mm thick, 50% glandular tissue). Images of the phantom were acquired using a BT prototype system that employs filtered backprojection for image reconstruction. The magnitude of the in-plane artifacts was evaluated and was found to increase linearly with increasing contrast (signal) level and size of the embedded objects. The contrast-detail phantom was found to be a useful tool for evaluating BT in-plane artifacts and might also be used to study out-of-plane artifacts and the effect of different acquisition and reconstruction parameters on image quality in BT.
© (2007) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Tony Svahn, Mark Ruschin, Bengt Hemdal, Lars Nyhlén, Ingvar Andersson, Pontus Timberg, Sören Mattsson, and Anders Tingberg "In-plane artifacts in breast tomosynthesis quantified with a novel contrast-detail phantom", Proc. SPIE 6510, Medical Imaging 2007: Physics of Medical Imaging, 65104R (16 March 2007); https://doi.org/10.1117/12.709591
Lens.org Logo
CITATIONS
Cited by 12 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical spheres

Breast

Polymethylmethacrylate

Signal attenuation

Image quality

Image quality standards

Mammography

Back to Top