You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
16 March 2007Extended volume coverage in helical cone-beam CT by using PI-line based BPF algorithm
We compared data requirements of filtered-backprojection (FBP) and backprojection-filtration (BPF) algorithms
based on PI-lines in helical cone-beam CT. Since the filtration process in FBP algorithm needs all the projection data
of PI-lines for each view, the required detector size should be bigger than the size that can cover Tam-Danielsson
(T-D) window to avoid data truncation. BPF algorithm, however, requires the projection data only within the T-D
window, which means smaller detector size can be used to reconstruct the same image than that in FBP. In other
words, a longer helical pitch can be obtained by using BPF algorithm without any truncation artifacts when a fixed
detector size is given. The purpose of the work is to demonstrate numerically that extended volume coverage in
helical cone-beam CT by using PI-line-based BPF algorithm can be achieved.
Seungryong Cho andXiaochuan Pan
"Extended volume coverage in helical cone-beam CT by using PI-line based BPF algorithm", Proc. SPIE 6510, Medical Imaging 2007: Physics of Medical Imaging, 651055 (16 March 2007); https://doi.org/10.1117/12.713237