You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
20 March 2007Technology assessment: observer study directly compares screen/film to CR mammography
A new study supports and expands upon a previous reporting that computed radiography (CR) mammography offers as
good, or better, image quality than state-of-the-art screen/film mammography. The suitability of CR mammography is
explored through qualitative and quantitative study components: feature comparison and cancer detection rates of each
modality. Images were collected from 150 normal and 50 biopsy-confirmed subjects representing a range of breast and
pathology types. Comparison views were collected without releasing compression, using automatic exposure control on
Kodak MIN-R films, followed by CR. Digital images were displayed as both softcopy (S/C) and hardcopy (H/C) for the
feature comparison, and S/C for the cancer detection task. The qualitative assessment used preference scores from five
board-certified radiologists obtained while viewing 100 screen/film-CR pairs from the cancer subjects for S/C and H/C
CR output. Fifteen general image-quality features were rated, and up to 12 additional features were rated for each pair,
based on the pathology present. Results demonstrate that CR is equivalent or preferred to conventional mammography
for overall image quality (89% S/C, 95% H/C), image contrast (95% S/C, 98% H/C), sharpness (86% S/C, 93% H/C),
and noise (94% S/C, 91% H/C). The quantitative objective was satisfied by asking 10 board-certified radiologists to
provide a BI-RADSTM score and probability of malignancy per breast for each modality of the 200 cases. At least 28
days passed between observations of the same case. Average sensitivity and specificity was 0.89 and 0.82 for CR and
0.91 and 0.82 for screen/film, respectively.
The alert did not successfully save. Please try again later.
Lynn Fletcher-Heath, Anne Richards, Susan Ryan-Kron, "Technology assessment: observer study directly compares screen/film to CR mammography," Proc. SPIE 6515, Medical Imaging 2007: Image Perception, Observer Performance, and Technology Assessment, 65151F (20 March 2007); https://doi.org/10.1117/12.712702