You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
26 April 2007The influence of human eye aberrations on the resolution and field of view of fundus-cameras
In this paper we estimate isoplanatic patch size of human eye using experimental results of human eye aberrations
measurements. For examined subjects it was found to be in the range 1.1° to 2.5°. The size of isoplanatic patch for
Gulistrand-Navarro model was calculated, it was found to be close to values obtained experimentally. We also measured
contribution of corneal surface and internal optics into total aberrations of the eye. We modified Gullstrand-Navarro eye
model to reproduce on-axis and off-axis performance of the eyes of each measured subject. The distribution of aberrations
between optical elements of the eye was taken into account when modeling. We also investigated isoplanatic patch size
widening methods such as average phase correction using 2 beacons and immersion method based on compensation of
external corneal surface refraction with immersion liquid. Immersion method was found to be the most appropriate for
isoplanatic patch widening as it allows us to increase isoplanatic patch size almost twice without loss of image quality at the
center of the image.
The alert did not successfully save. Please try again later.
Alexander Dubinin, Tatyana Cherezova, Alexey Belyakov, Alexis Kudryashov, "The influence of human eye aberrations on the resolution and field of view of fundus-cameras," Proc. SPIE 6535, Saratov Fall Meeting 2006: Optical Technologies in Biophysics and Medicine VIII, 653506 (26 April 2007); https://doi.org/10.1117/12.740606