Paper
13 July 2007 Partial kidney resection based on 1.94μm fiber laser system
Author Affiliations +
Abstract
A 1.94 μm Laser-Scalpel system (Starmedtec, Starnberg, Germany) emitting at a wavelength at 1.94 μm and a max. laser power 18 W (cw mode) was used for partial resection of porcine kidney. Partial resection was performed on 9 kidneys. Six kidneys were extracted directly after resection and pigs were sacrificed. The remaining three pigs were kept alive for 2 weeks post resection in order to investigate the healing process. No complications during the operation and during the 2 weeks survival period were found. Total resection time including haemostasis of the remaining tissue was 10.2±6.5 min, blood lost was 14.7±29 ml. Haemostatic property of the 1.94 μm Laser-Scalpel was sufficient. Histological evaluation with H&E stained tissue samples showed medium carbonization and a thermal damage zone of 1mm. In conclusion, the first experiments show that the 1.94 μm Laser-Scalpel is a very promising device for bloodless and fast kidney resection.
© (2007) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
D. Theisen-Kunde, S. Tedsen, K. Herrmann, V. Danicke, and R. Brinkmann "Partial kidney resection based on 1.94μm fiber laser system", Proc. SPIE 6632, Therapeutic Laser Applications and Laser-Tissue Interactions III, 663205 (13 July 2007); https://doi.org/10.1117/12.727861
Lens.org Logo
CITATIONS
Cited by 5 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Kidney

Tissues

Fiber lasers

Laser systems engineering

Laser cutting

Continuous wave operation

Quartz

Back to Top