21 September 2007 Performance measurements of Al/CdTe/Pt pixel diode detectors
Author Affiliations +
Abstract
We developed Schottky CdTe detectors using Al as an anode electrode and measured their performances. We first fabricated monolithic detectors with four different thicknesses of 0.5, 0.75, 1.0, and 2.0 mm. An Al anode electrode was implemented with a guard-ring structure. For the 0.5 mm thick CdTe detector, an energy resolution of 1.2 keV (FWHM) at 122 keV was achieved at a temperature of −20 °C and a bias voltage of 400 V. Using the same technology, we next developed 8 × 8 pixel CdTe detectors, again with the four different thicknesses. The Al anode electrode was pixelated and the Pt cathode was made as a single plate. Signals from all pixels were successfully obtained and an energy resolution of 1.3 keV and 1.9 keV (FWHM) for 59.5 keV and 122 keV gamma-rays, was achieved at a temperature of −20 °C and a bias voltage of 400 V using the 0.5 mm thick CdTe detector. The energy resolution was nearly the same in each pixel.
© (2007) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Shin-nosuke Ishikawa, Hiroyuki Aono, Shin Watanabe, Shin'ichiro Takeda, Kazuhiro Nakazawa, and Tadayuki Takahashi "Performance measurements of Al/CdTe/Pt pixel diode detectors", Proc. SPIE 6706, Hard X-Ray and Gamma-Ray Detector Physics IX, 67060M (21 September 2007); https://doi.org/10.1117/12.735188
Lens.org Logo
CITATIONS
Cited by 7 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Electrodes

Diodes

Aluminum

Gamma radiation

Image resolution

Platinum

Back to Top