Paper
28 November 2007 Error analysis for integrating sphere of color measuring system
Author Affiliations +
Abstract
Integrating sphere has been widely used in the photoelectric integrating color measurement. Ordinarily, the measurement caliber of integrating sphere is smaller than the diameter of incidence beam, and some of the incidence beam irradiate on the inner surface of integrating sphere directly. It introduces error into the color measurement. The paper establishes the mathematics model of the error introduced by integrating sphere in color measurement, and has found the way to revise the error. The result indicates that this method can revise the error which is carried by the mismatch between the measurement caliber of integrating sphere and the diameter of incidence beam effectively.
© (2007) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Shang-zhong Jin and Kun Yuan "Error analysis for integrating sphere of color measuring system", Proc. SPIE 6834, Optical Design and Testing III, 68343U (28 November 2007); https://doi.org/10.1117/12.759230
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Integrating spheres

Light sources

Sensors

Calibration

Black bodies

Reflectivity

Error analysis