There have been several technologies to enable high resolution cross-sectional images of biological tissues in optical coherence tomography (OCT) method. Optical frequency comb (OFC) source has been proposed to overcome the crosstalk problem among the CCD detector pixels of the continuous spectrum of light source. Recently, a passive-type OFC is demonstrated simply placing a Fabry-Perot interferometer filter right after the broadband light source, but it shows a high loss of output light power and limited tenability of channel spacing of multi-wavelength. In this work, we experimentally demonstrate a spectral comparison of a novel multi-wavelength source based on a fiber Sagnac interferometer. The channel spacing is flexibly tuned by the effective length control of polarization-maintaining fiber (PMF). The uniform and stable multi-wavelength spectral distribution is also helpful to obtain the higher sensitivity from the lower exposure intensity source to get a better quality spectral OCT image.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Eun Joo Jung, Jae-Seok Park, Myung Yung Jeong, Chang-Seok Kim, Tae Joong Eom, Vitali A. Tougbaev, Bong-Ahn Yu, Woojin Shin, Do-Kyeong Ko, and Ji-Hye Lee "Multi-wavelength source for the lower exposure intensity of spectral OCT", Proc. SPIE 6849, Design and Quality for Biomedical Technologies, 68490A (12 February 2008); https://doi.org/10.1117/12.762917
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical coherence tomography

Light sources

Charge-coupled devices

Optical filters

Signal to noise ratio

Glasses

Point spread functions

Back to Top