You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
12 August 1986Characterization Of The Absolute Photon Sensitivity Of Gold Cathode Photoelectric Detectors
The responses of gold cathode photoelectric detectors to photons in the 3 to 7.5 keV energy band have been measured using synchrotron radiation. The photoelectron detectors under study have cathodes with a conical configurations. Their sensitivities with respect to position on the cathode are nearly uniform (within 30 percent) over approximately 65 percent of the cone surface. The responses fall rapidly near the outer edges. Sensitivities with respect to the photon angle of incidence are proportional to (sin θ)-1. In the photon energy band covered, the measured response of the detectors are approximately 20 percent lower than the responses measured with conventional x-ray sources. These differences are due to differences in the geometries of the two calibrations.
The alert did not successfully save. Please try again later.
C. M. Dozier, R. K. Freitag, D. L. Fehl, "Characterization Of The Absolute Photon Sensitivity Of Gold Cathode Photoelectric Detectors," Proc. SPIE 0689, X-Ray Calibration: Techniques, Sources, and Detectors, (12 August 1986); https://doi.org/10.1117/12.936569