You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
17 March 2008Radiation therapy planning and simulation with magnetic resonance images
We present a system which allows for use of magnetic resonance (MR) images as primary RT workflow modality alone
and no longer limits the user to computed tomography data for radiation therapy (RT) planning, simulation and patient
localization. The single steps for achieving this goal are explained in detail.
For planning two MR data sets, MR1 and MR2 are acquired sequentially. For MR1 a standardized Ultrashort TE (UTE)
sequence is used enhancing bony anatomy. The sequence for MR2 is chosen to get optimal contrast for the target and the
organs at risk for each individual patient. Both images are naturally in registration, neglecting elastic soft tissue
deformations. The planning software first automatically extracts skin and bony anatomy from MR1. The user can semi-automatically
delineate target structures and organs at risk based on MR1 or MR2, associate all segmentations with MR1
and create a plan in the coordinate system of MR1. Projections similar to digitally reconstructed radiographs (DRR)
enhancing bony anatomy are calculated from the MR1 directly and can be used for iso-center definition and setup
verification. Furthermore we present a method for creating a Pseudo-CT data set which assigns electron densities to the
voxels of MR1 based on the skin and bone segmentations. The Pseudo-CT is then used for dose calculation.
Results from first tests under clinical conditions show the feasibility of the completely MR based workflow in RT for
necessary clinical cases. It needs to be investigated in how far geometrical distortions influence accuracy of MR-based
RT planning.
The alert did not successfully save. Please try again later.
Thomas Boettger, Tufve Nyholm, Magnus Karlsson, Chandrasekhar Nunna, Juan Carlos Celi, "Radiation therapy planning and simulation with magnetic resonance images," Proc. SPIE 6918, Medical Imaging 2008: Visualization, Image-Guided Procedures, and Modeling, 69181C (17 March 2008); https://doi.org/10.1117/12.770016