You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
29 April 2008High-energy (MeV) x-ray imaging with a mercuric iodide imager
A 13 cm × 13 cm size mercuric iodide (HgI2) imager was tested for high (MeV) energy X-ray imaging with a 6 MeV
Betatron and a 4 MeV Linac x-ray source, and the data were compared to those measured at diagnostic (keV) energies.
The 127 μm pixel size imager gave excellent resolution, with a MTF (modulation transfer function) of 45% at the
Nyquist frequency, similar to that of measured at diagnostic (keV) energies. The sensitivity of the imager was measured
using a 1 mm thick Cu and a 0.5 mm thick Ta buildup plate, placed on the top electrode of the HgI2 layer and also
without any buildup plate. The highest signal levels were obtained without a buildup plate.
The imager can also capture fluoroscopic images at up to 15 fr/sec in the full resolution mode, and at up to 30 fr/sec
with 2 × 2 pixel binning. . With this imager, small steel objects were clearly visible behind a 1/8" thick steel plate.
These experiments with high energy x-rays, demonstrate that HgI2 imagers can be used, not only for the diagnostic
energy range, but also for the MeV energy range. Moreover the same imager can be used for dual energy (keV and
MeV) imaging for medical, NDE (non destructive evaluation) and homeland security applications.
G. Zentai andL. Partain
"High-energy (MeV) x-ray imaging with a mercuric iodide imager", Proc. SPIE 6935, Health Monitoring of Structural and Biological Systems 2008, 69351Y (29 April 2008); https://doi.org/10.1117/12.776537