Paper
16 May 2008 Respiratory gating of endoscopic OCT images of the upper airway
Robert A. McLaughlin, Julian J. Armstrong, Sven Becker, Jennifer H. Walsh, Jason Kirkness, Arpit Jain, Matthew S. Leigh, Jonathan Williamson, David R. Hillman M.D., Peter R. Eastwood, David D. Sampson
Author Affiliations +
Proceedings Volume 7004, 19th International Conference on Optical Fibre Sensors; 70045V (2008) https://doi.org/10.1117/12.786827
Event: 19th International Conference on Optical Fibre Sensors, 2008, Perth, WA, Australia
Abstract
Anatomical optical coherence tomography (aOCT) is an endoscopic imaging modality that can be used to quantify size and shape of the upper airway. We report the application of respiratory gating to aOCT images. Our results show that respiratory gating can reduce motion artefact in upper airway images. Using an error metric based on distance to the dominant reflection in each A-scan, we found notable improvements when the breath cycle was partitioned into approximately four gates, but only minor improvements as the number of gates was further increased.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Robert A. McLaughlin, Julian J. Armstrong, Sven Becker, Jennifer H. Walsh, Jason Kirkness, Arpit Jain, Matthew S. Leigh, Jonathan Williamson, David R. Hillman M.D., Peter R. Eastwood, and David D. Sampson "Respiratory gating of endoscopic OCT images of the upper airway", Proc. SPIE 7004, 19th International Conference on Optical Fibre Sensors, 70045V (16 May 2008); https://doi.org/10.1117/12.786827
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical coherence tomography

Endoscopy

Lung

Data acquisition

Distance measurement

Error analysis

Head

Back to Top