You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
22 July 2008VIRUS-P: camera design and performance
We present the design and performance of the prototype Visible Integral-field Replicable Unit Spectrograph
(VIRUS-P) camera. Commissioned in 2007, VIRUS-P is the prototype for 150+ identical fiber-fed integral field
spectrographs for the Hobby-Eberly Telescope Dark Energy Experiment. With minimal complexity, the gimbal
mounted, double-Schmidt design achieves high on-sky throughput, image quality, contrast, and stability with novel
optics, coatings, baffling, and minimization of obscuration. The system corrector working for both the collimator
and f / 1.33 vacuum Schmidt camera serves as the cryostat window while a 49 mm square aspheric field flattener sets
the central obscuration. The mount, electronics, and cooling of the 2k × 2k, Fairchild Imaging CCD3041-BI fit in
the field-flattener footprint. Ultra-black knife edge baffles at the corrector, spider, and adjustable mirror, and a
detector mask, match the optical footprints at each location and help maximize the 94% contrast between 245
spectra. An optimally stiff and light symmetric four vane stainless steel spider supports the CCD which is thermally
isolated with an equally stiff Ultem-1000 structure. The detector/field flattener spacing is maintained to 1 μm for all
camera orientations and repeatably reassembled to 12 μm. Invar rods in tension hold the camera focus to ±4 μm
over a -5-25 °C temperature range. Delivering a read noise of 4.2 e- RMS, sCTE of 1-10-5 , and pCTE of
1-10-6 at
100 kpix/s, the McDonald V2 controller also helps to achieve a 38 hr hold time with 3 L of LN2 while maintaining
the detector temperature setpoint to 150 μK (5σ RMS).