You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
22 July 2008Count rate dependent non-linearity and pixel size variations in 1.7 micron cut-off detectors
Over the past decade scientists have collected convincing evidence that the content of our universe is dominated
by a mysterious dark energy. Understanding the nature of dark energy is a very difficult task, and requires a
variety of independent experimental approaches. Most of these approaches rely on photometric calibrations over
a wide range of intensities using standardized stars and internal reference sources, and hence on a complete
understanding of the linearity of the detectors. The SNAP near infrared (NIR) instrument team has performed
a comprehensive study of precision photometry on 1.7 micron cut-off HgCdTe detectors. Among those studies
are the count rate dependent detector non-linearity that was recently discovered with the NICMOS array on
the Hubble Space Telescope, and possible pixel size variations seen in quantum efficiency (QE) data. The nonlinearity
on NICMOS exhibits an unexpected behavior, where pixels with high (low) count rates detect slightly
more (less) flux than expected for a linear system. To test this count rate dependent non-linearity a dedicated
setup was built that produces a known amount of light on a detector, and measures its response as a function of
light intensity and wavelength. If the pixel response variations seen in QE data are due to pixel area variations,
standard flat-fielding will degrade photometry precision for point sources in an undersampled telescope. Studies
have been performed to estimate the magnitude of pixel area variations.
The alert did not successfully save. Please try again later.
Wolfgang Lorenzon, Robert Newman, Michael Schubnell, G. Tarlé, C. Weaverdyck, "Count rate dependent non-linearity and pixel size variations in 1.7 micron cut-off detectors," Proc. SPIE 7021, High Energy, Optical, and Infrared Detectors for Astronomy III, 70210V (22 July 2008); https://doi.org/10.1117/12.789638