13 March 2009 Cardiac imaging with multi-sector data acquisition in volumetric CT: variation of effective temporal resolution and its potential clinical consequences
Xiangyang Tang, Jiang Hsieh, Basel H. Taha, Melissa L. Vass, John L. Seamans, Darin R. Okerlund
Author Affiliations +
Proceedings Volume 7258, Medical Imaging 2009: Physics of Medical Imaging; 725832 (2009) https://doi.org/10.1117/12.812231
Event: SPIE Medical Imaging, 2009, Lake Buena Vista (Orlando Area), Florida, United States
Abstract
With increasing longitudinal detector dimension available in diagnostic volumetric CT, step-and-shoot scan is becoming popular for cardiac imaging. In comparison to helical scan, step-and-shoot scan decouples patient table movement from cardiac gating/triggering, which facilitates the cardiac imaging via multi-sector data acquisition, as well as the administration of inter-cycle heart beat variation (arrhythmia) and radiation dose efficiency. Ideally, a multi-sector data acquisition can improve temporal resolution at a factor the same as the number of sectors (best scenario). In reality, however, the effective temporal resolution is jointly determined by gantry rotation speed and patient heart beat rate, which may significantly lower than the ideal or no improvement (worst scenario). Hence, it is clinically relevant to investigate the behavior of effective temporal resolution in cardiac imaging with multi-sector data acquisition. In this study, a 5-second cine scan of a porcine heart, which cascades 6 porcine cardiac cycles, is acquired. In addition to theoretical analysis and motion phantom study, the clinical consequences due to the effective temporal resolution variation are evaluated qualitative or quantitatively. By employing a 2-sector image reconstruction strategy, a total of 15 (the permutation of P(6, 2)) cases between the best and worst scenarios are studied, providing informative guidance for the design and optimization of CT cardiac imaging in volumetric CT with multi-sector data acquisition.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Xiangyang Tang, Jiang Hsieh, Basel H. Taha, Melissa L. Vass, John L. Seamans, and Darin R. Okerlund "Cardiac imaging with multi-sector data acquisition in volumetric CT: variation of effective temporal resolution and its potential clinical consequences", Proc. SPIE 7258, Medical Imaging 2009: Physics of Medical Imaging, 725832 (13 March 2009); https://doi.org/10.1117/12.812231
Lens.org Logo
CITATIONS
Cited by 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Temporal resolution

Data acquisition

Computed tomography

Cardiac imaging

Heart

Image restoration

Sensors

Back to Top