You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
13 March 2009Dose efficiency consideration for volume-of-interest breast imaging using x-ray differential phase-contrast CT
The newly developed differential phase-contrast (DPC) imaging technique has attracted increasing interest among
researchers. In a DPC system, the self-imaging effect and the
phase-stepping method are implemented through three
gratings to manifest phase contrast, and differentiated phase images can be obtained. An important advantage of this
technique is that hospital-grade x-ray tubes can be used, allowing much higher x-ray output power and faster image
processing than with micro-focus in-line phase-contrast imaging. A DPC-CT system can acquire images from different
view angles along a circular orbit, and tomographic images can be reconstructed. However, the principle of DPC
imaging requires multiple exposures to compute any differentiated phase image at each view angle, which raises
concerns about radiation exposure via x-ray dose. Computer simulations are carried out to study the dose efficiency for
DPC-CT for volume-of-interest breast imaging. A conceptual
CBCT/DPC-CT hybrid imaging system and a numerical
breast phantom are designed for this study. A FBP-type reconstruction algorithm is optimized for the VOI
reconstruction. Factors including the x-ray flux and detector pixel size are considered and their effects on reconstruction
image quality in terms of noise level and contrast-to-noise ratio are investigated. The results indicate that with a pixel
size of 20 microns and a dose level of 5.7mGy, which is equivalent to the patient dose of a two-view mammography
screening or a dedicated CBCT breast imaging scan, much better tissue contrast and spatial resolution can be achieved
using the DPC-CT technique. It is very promising for possible application at pathology-level in vivo study for human
breasts.