You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
18 March 201050 μm pixel size a-Se mammography imager with high DQE and increased temperature resistance
The imager presented in this paper has a special blocking structure that ensures very low dark current of less than 1
pA/mm2 even with a 20 V/μm electric field. Hence the electric field can be increased from the generally applied 10
V/μm to 20V/μm, this reduces the energy required to produce an electron hole (e-h) pair from 60 eV to about 36 eV at
the given (19.3 keV mean) mammo energy. Furthermore, with special doping and manufacturing processes this a-Se
layer is very stable in the 0-70 C° temperature range as demonstrated by Ogusu et al. [1].
A new 5 cm × 5 cm size TFT array was developed with 50 μm pixel size, specifically for testing the resolution of
photoconductor based imagers. The first new imager of this type had a 200 μm thick a-Se layer evaporated onto the
array. Its MTF, NPS, and DQE values were evaluated using 28kVp Mo anode x-ray source with a 0.03mm thick Mo and
an additional 2 mm thick Al filters. The MTF value is about 40% and 50% in x-and y directions at the Nyquist frequency
of 10 lp/mm. The low frequency DQE at 20 V/μm electrical field is ~70% at 151 μGy dose and drops only about 10%
when going down to 23 μGy.
This new array also has excellent lag properties. The measured first frame image lag at 20 V/μm is less than 1%. Such
low lag provides opportunities to use this material not only for mammography but also for breast tomosynthesis
applications. Breast phantom images demonstrate that even the smallest 0.13 mm calcifications are clearly visible with
this high-resolution imager.