22 March 2010 Designing a phantom for dose evaluation in multi-slice CT
Samir Abboud, Andreu Badal, Stanley H. Stern, Iacovos S. Kyprianou
Author Affiliations +
Abstract
Accurately representing radiation dose delivered in MSCT is becoming a concern as the maximum beam width of some modern CT scanners tends to become wider than the 100 mm charge-collection length of the pencil ionization chamber generally used in CT dosimetry. We investigate two alternative methods of dose evaluation in CT scanners. We investigate two alternative approaches for better characterization of CT dose than conventional evaluation of CTDI100. First, we simulate dose profiles and energy deposition in phantoms longer than the typically used 14-15 cm length right-circular cylinders. Second we explore the accuracy and practicality of applying mathematical convolution to a scatter kernel in order to generate dose profiles. A basic requirement for any newly designed phantom is that it be able to capture approximately the same dose as would an infinitely long cylinder, but yet be of a size and weight that a person could easily carry and position. Using the PENELOPE Monte Carlo package, we simulated dose profiles in cylindrical polymethyl methacrylate (PMMA) phantoms of 10, 16, 20, 24 and 32 cm diameter and 15, 30 and 300 cm length. Beam widths were varied from 1 cm to 60 cm. Lengths necessary to include within the dose integrals values associated with the scatter tails as well as with the primary radiation of the profile were then calculated as the full width at five percent of maximum dose. The resulting lengths suggest that to accommodate wide beam widths, phantoms longer than those currently used are necessary. The results also suggest that using a longer phantom is a relatively more accurate approach, while using mathematical convolution is simpler and more practical to implement than using the long phantoms designed according to direct Monte Carlo simulations.
© (2010) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Samir Abboud, Andreu Badal, Stanley H. Stern, and Iacovos S. Kyprianou "Designing a phantom for dose evaluation in multi-slice CT", Proc. SPIE 7622, Medical Imaging 2010: Physics of Medical Imaging, 762232 (22 March 2010); https://doi.org/10.1117/12.844355
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Monte Carlo methods

Convolution

Scanners

Computed tomography

Optical simulations

Mathematical modeling

Polymethylmethacrylate

Back to Top