You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
22 March 2010Off-center object of interest (OOI) imaging in filtered region of interest rotational angiography (FROI-RA)
Rotational angiography (RA) is widely used clinically to obtain 3D data. In many procedures, e.g.,
neurovascular interventions, the imaged field of view (FOV) is much larger than the region of interest (ROI),
thereby subjecting the patient to unnecessary x-ray dose. To reduce the dose in these procedures, we have proposed
placing an x-ray attenuating filter with an open aperture (ROI) in the x-ray beam (called filtered region of interest
(FROI) RA. We have shown that this approach yields high quality data for centered objects of interest (OoIs). In
this study, we investigate the noise behavior of the FROI approach for off-center OoIs.
Using filter-specific attenuation and noise characteristics, simulated FROI projection images were generated.
The intensities in the peripheral region were equalized, and the 3D data reconstructed. For each reconstructed voxel,
the intersections with the full intensity beam (ROI) were determined for each projection, and noise properties were
evaluated. Off-center OoIs intersect the high intensity beam in more than 60% of the projections (ROI having 40%
FOV area), with intersection frequency increasing with increasing ROI area and OoI proximity to the central region.
The noise increases with distance from the central region up to a factor of two. Integral dose reductions range
between 40% and 85%, depending on ROI area and filter thickness.
Substantial dose reductions (40-85%) are achieved with less than a factor of two increase in noise for OoIs
peripheral to the central region, indicating the FROI approach might be an alternative for reducing dose during
standard procedures.
The alert did not successfully save. Please try again later.
Sebastian Schafer, Peter B. Noël, Alan M. Walczak, Andrew Kuhls-Gilcrist, Kenneth R. Hoffmann, "Off-center object of interest (OOI) imaging in filtered region of interest rotational angiography (FROI-RA)," Proc. SPIE 7622, Medical Imaging 2010: Physics of Medical Imaging, 76223F (22 March 2010); https://doi.org/10.1117/12.844011