You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
23 February 2010Absolute vs. relative error characterization of
electromagnetic tracking accuracy
Electromagnetic (EM) tracking systems are often used for real time navigation of medical tools in an Image
Guided Therapy (IGT) system. They are specifically advantageous when the medical device requires tracking
within the body of a patient where line of sight constraints prevent the use of conventional optical tracking. EM
tracking systems are however very sensitive to electromagnetic field distortions. These distortions, arising from
changes in the electromagnetic environment due to the presence of conductive ferromagnetic surgical tools or
other medical equipment, limit the accuracy of EM tracking, in some cases potentially rendering tracking data
unusable. We present a mapping method for the operating region over which EM tracking sensors are used,
allowing for characterization of measurement errors, in turn providing physicians with visual feedback about
measurement confidence or reliability of localization estimates.
In this instance, we employ a calibration phantom to assess distortion within the operating field of the
EM tracker and to display in real time the distribution of measurement errors, as well as the location and
extent of the field associated with minimal spatial distortion. The accuracy is assessed relative to successive
measurements. Error is computed for a reference point and consecutive measurement errors are displayed relative
to the reference in order to characterize the accuracy in near-real-time. In an initial set-up phase, the phantom
geometry is calibrated by registering the data from a multitude of EM sensors in a non-ferromagnetic ("clean")
EM environment. The registration results in the locations of sensors with respect to each other and defines
the geometry of the sensors in the phantom. In a measurement phase, the position and orientation data from
all sensors are compared with the known geometry of the sensor spacing, and localization errors (displacement
and orientation) are computed. Based on error thresholds provided by the operator, the spatial distribution of
localization errors are clustered and dynamically displayed as separate confidence zones within the operating
region of the EM tracker space.
The alert did not successfully save. Please try again later.
Mohammad Matinfar, Ganesh Narayanasamy, Luis Gutierrez, Raymond Chan, Ameet Jain, "Absolute vs. relative error characterization of electromagnetic tracking accuracy," Proc. SPIE 7625, Medical Imaging 2010: Visualization, Image-Guided Procedures, and Modeling, 762524 (23 February 2010); https://doi.org/10.1117/12.844326