You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
18 March 2010Dynamic CT head phantom for perfusion and angiography studies
Contrast imaging is a compelling enhancement for the portable, flat panel-based brain CT scanner currently under
development at Xoran. Due to the relative low temporal resolution of flat panel detectors, enabling tomographic imaging
on such platform requires optimizing the imaging and injection protocols. A dynamic CT head phantom was designed to
facilitate this task. The Dynamic Perfusion and Angiography Model (PAM), mimics tissue attenuation in CT images,
provides physiological timing for angiography and perfusion studies, and moves fluid with properties similar to those of
blood. The design consists of an arterial system, which contains bifurcating vessels that feed into perfusion chambers,
mimicking blood flow through capillaries and smaller vessels, and a venous system, which is symmetrical to the arterial
side and drains the perfusion chambers. The variation of geometry and flow rate in the phantom provides the
physiological total time that fluid spends in the head, and the difference in material densities correlates to CT numbers
for biological tissues. This paper discusses the design of Dynamic PAM and shows experimental results demonstrating
its ability to realistically simulate blood flow. Results of dynamic imaging studies of the phantom are also presented.
The alert did not successfully save. Please try again later.
K. Russell, A. Blazeski, K. Dannecker, Q. Y. Lee, C. Holscher, C. Donahue, W. van Kampen, "Dynamic CT head phantom for perfusion and angiography studies," Proc. SPIE 7626, Medical Imaging 2010: Biomedical Applications in Molecular, Structural, and Functional Imaging, 76261Y (18 March 2010); https://doi.org/10.1117/12.844474