You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
9 November 2010Design of imaging spectrometer based on Czerny-Turner in FUV
It is the astigmatism that leads the traditional imaging spectrometer based on Czerny-Turner to have low spatial
resolution. And it is discovered that when the distance between concave mirror and grating, x, is equal to the twice of
focal length, , of the mirror, SII = SIII = 0 and the aberration is the least as well as the astigmatism is eliminated greatly.
Meanwhile the toroidal mirror is presented to correct the astigmatism, and as well the aberration caused by the large
FOV is corrected by optimizing the surface tilt. Then both of the spatial and spectral resolutions are improved. Finally a
Czerny-Turner imaging spectrometer working in FUV (120 nm ~ 180 nm) with 2.5° FOV is designed, and its focal
length is 147.61 mm, its F number is 3.93. MTF of this imaging spectrometer is more than 0.39 at 20 lp/mm in the total
wavelength band of FOV, which satisfied the requirements of imaging spectrometer working on satellite in FUV.