You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
11 February 2011Critical evaluation of the ultrasonic pachymetry for "in vitro" corneas
The measurement of central corneal thickness (CCT) is vastly useful for diagnostic and therapeutic evaluation. The
ultrasound pachymetry is currently the most common CCT technique. This study was undertaken to determine the
precision and correlation of measurements obtained by mechanical and ultrasound pachymetry. The ultrasound
pachymetry was determined using an A-scan ultrasonic pachymeter. The probe tip was held perpendicular on the central
cornea by a support that goes down smooth to avoid excessive pressure and instability. The mechanical pachymetry was
determined using a micrometer with a tip of 2mm of diameter. The tip of the micrometer was held perpendicular on the
central cornea by a support that keeps stabilized. A 10x optics increase and a digital video camera shows real time image
of approach and full contact of the tip with the cornea. Eight human corneas were obtained from cadaveric eyes.
Measurements in both systens were taken for three different users, each one performed five readings. The results for both
systems has an average SD of 33 microns refers to the systematic error between users (for positioning, center, pinching).
But the difference between systems was 120 microns, possibly refers to the imprecision of ultrasound pachymetry in
measuring in vitro corneas.