You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
10 February 2011System-independent assessment of OCT axial resolution with a "bar
chart" phantom
We present a optical phantom approach for the characterization of OCT axial resolution and contrast via multilayered
"bar charts." We explored two methods to fabricate these phantoms: the first is based on monolayers of light-scattering
microspheres with an intervening layer of transparent silicone, and the second involves alternating layers of scatteringenhanced
silicone and transparent silicone. Varying the diameter of the microspheres and the thickness of the silicone
layers permits different spatial frequencies to be realized in the axial dimension of the phantoms. Because the phantom's
dimensions are accurately known independent of the OCT system, no information about the system's spatial calibration
is required. We quantified the degree to which the bars in each phantom could be resolved with OCT, which provides
insights into the axial contrast transfer function. Initial testing of these phantoms on time-resolved and Fourier-domain
OCT platforms indicated that our approach can provide accurate, system-independent estimates of resolution.
The alert did not successfully save. Please try again later.
Anant Agrawal, Robert Chang, Megan Connors, Christopher Stafford, Jeeseong Hwang, Joshua Pfefer, "System-independent assessment of OCT axial resolution with a "bar chart" phantom," Proc. SPIE 7906, Optical Diagnostics and Sensing XI: Toward Point-of-Care Diagnostics; and Design and Performance Validation of Phantoms Used in Conjunction with Optical Measurement of Tissue III, 79060R (10 February 2011); https://doi.org/10.1117/12.880958