You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
16 March 20113D task-based performance assessment metrics for optimization
of performance and dose in breast tomosynthesis
This study aimed to investigate a method for empirically evaluating 3D imaging task performance of breast
tomosynthesis imaging systems. A simulation and experimental approach was used to develop a robust
method for performance assessment. To identify a method for experimentally assessing the 3D modulation
transfer function (MTF), a breast tomosysnthesis system was first simulated using cascaded system analysis
to model the signal and noise characteristics of the projections. A range of spheres with varying contrast and
size were reconstructed using filtered back projection from which the 3D MTF was evaluated. Results
revealed that smaller spheres result in lower artifacts in the measured MTF, where a sphere of 0.5 mm was
found ideal for experimental purposes. A clinical tomosynthesis unit was used as a platform for quantifying
the effect of acquisition and processing parameters (e.g., angular extent and sampling, dose, and voxel size)
on breast imaging performance. The 3D noise-power spectrum (NPS) was measured using a uniform phantom
and 3D MTF was measured using 0.5 mm ruby spheres. These metrics were combined with a mathematical
description of imaging task to generate a figure of merit called the detectability index for system evaluation
and optimization. Clinically relevant imaging tasks were considered, such as the detection and localization of
a spherical mass. The detectability index was found to provide a useful metric that accounts for the complex
3D imaging characteristics of breast tomosynthesis. Results highlighted the dependence of optimal technique
on the imaging task. They further provided initial validation of an empirically assessed figure of merit for
clinical performance assessment and optimization of breast tomosynthesis systems.
S. Richard andE. Samei
"3D task-based performance assessment metrics for optimization
of performance and dose in breast tomosynthesis", Proc. SPIE 7961, Medical Imaging 2011: Physics of Medical Imaging, 79611A (16 March 2011); https://doi.org/10.1117/12.877070