You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
15 September 2011Optimizing CdTe detectors and ASIC readouts for high-flux x-ray imaging
Developments in room temperature cadmium telluride (CdTe) based solid state imaging arrays for energy-resolved
photon-counting in medical x-ray imaging are discussed. A number of x-ray imaging applications can benefit from these
developments including mammography, radiography, and computed tomography (CT). Energy-resolved photon-counting
can provide reduced dose through optimal energy weighting, compositional analysis through multiple basis
function material decomposition, and contrast enhancement through spectroscopic x-ray imaging of metal nanoparticles.
Extremely high flux can occur in x-ray imaging and energy integrating detectors have been conventionally used. To
achieve the benefits of energy resolved photon counting, imaging arrays with a large count rate range and good detection
efficiency are required. Compound semiconductor radiation detectors with pixellated anode arrays electrically connected
to application specific integrated circuits (ASICs) can provide fast, efficient, low-noise performance with adequate
energy resolution however this can only be achieved with a careful optimization of the CdTe sensors and ASICs
together. We have designed and constructed CdTe imaging arrays, 3 mm thick with a grid of electrical contacts inter-connected
to a multi-channel channel ASICs. Arrays with a pixel pitch of 0.5 mm have achieved a counting range up to
20 million counts per second per square mm. Additionally, ASICs with a two dimensional array of pads has been
fabricated and tested by connecting the inputs to 1 mm pitch CdTe sensors demonstrating 7 keV full width at half
maximum energy resolution across a dynamic range of 30 keV to 140 keV for clinical CT.
The alert did not successfully save. Please try again later.
William C. Barber, Einar Nygard, Jan C. Wessel, Nail Malakhov, Neal E. Hartsough, Thulasi Ghandi, Gregor Wawrzyniak, Jan S. Iwanczyk, "Optimizing CdTe detectors and ASIC readouts for high-flux x-ray imaging," Proc. SPIE 8143, Medical Applications of Radiation Detectors, 81430J (15 September 2011); https://doi.org/10.1117/12.897493