You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
21 September 2012Count rate nonlinearity in near infrared detectors: inverse persistence
The count rate non-linearity of near-infrared devices was first found in the HST NICMOS. In this report we present a
physical model of the cause of this effect, show how it is related to persistence, and compare the predictions of the model
to other observations of anomalous detector behavior. This model is able to explain not only the count rate non-linearity
but also several other effects. Overall, the excellent agreement between this model and the observations gives us strong
confidence that we understand the underlying cause of the count rate non-linearity. This understanding should allow us
to develop methods to accurately calibrate and remove the effect from JWST observations.