You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
17 October 2012Energy-resolved Compton scatter estimation for micro-CT
X-ray scatter can cause significant distortion in CT imaging, especially with the move to cone-beam geometries. Incoherent scatter (Compton scatter) is known to reduce the energy of scattered photons according to the angle of the scattering. The emergence of energy-resolved x-ray detectors offers an opportunity to produce and apply more accurate scatter estimates, leading to improved image quality. We have developed a scatter estimation algorithm that accounts for the variation in scatter with incident radiation energy. Where existing methods generate estimates of scatter for the complete detected energy band, our new method produces separate estimates for each of the energy bands that are measured, allowing a more focused correction of scatter. Our method is intended to be used in an iterative compensation framework like that of Rührnschopf and Klingenbeck (2011); it calculates the scatter contribution to each energy bin used in a scan based on the current volume estimate. Comparisons with Monte Carlo simulations indicate that this algorithm is effective at estimating the scatter level in separate energy bins. We found that the amount of scatter that loses enough energy to hop between energy bands is small enough to neglect, but that scatter intensity is dependent on the incident energy, so application of a spectrally-aware compensation technique is valuable.
The alert did not successfully save. Please try again later.
A. M. T. Opie, A. P. H. Butler, P. J. Bones, "Energy-resolved Compton scatter estimation for micro-CT," Proc. SPIE 8506, Developments in X-Ray Tomography VIII, 850616 (17 October 2012); https://doi.org/10.1117/12.930319