You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
23 October 2012Polarization sensitivity of Visible Infrared Imager Radiometer Suite (VIIRS) Flight Unit 1
This paper describes measured performance of VIIRS Flight unit 1 (F1) in response to variations in linearly polarized light
input into the instrument. Measurements and analysis show that VIIRS F1 meets specified polarization sensitivity
requirements with maximum measured degree of linear polarization (DoLP) ranging from 0.5% for the M7 band (865 nm) to
2.8% for the M1 band at 412 nm. Estimated uncertainty in the DoLP ranges from 0.l7% for the M6 band at 746 nm up to
0.37% for I2 at 865 nm. Detailed tables of measured polarization sensitivity characteristics for all spectral bands, all detectors
and both sides of the VIIRS de-rotation mirror are provided to help the VIIRS users’ community with analysis of measured
spectral radiances in the solar reflectance channels.
The alert did not successfully save. Please try again later.
Edward J. Novitsky, Stephen Herbst, James B. Young, Jeffery J. Puschell, Eric C. Fest, Bruce Guenther, "Polarization sensitivity of Visible Infrared Imager Radiometer Suite (VIIRS) Flight Unit 1," Proc. SPIE 8516, Remote Sensing System Engineering IV, 85160I (23 October 2012); https://doi.org/10.1117/12.2000151