You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
19 March 2013Spatial resolution characteristics of a-Se imaging detectors using
Monte Carlo methods with detailed spatiotemporal transport of x-rays,
electrons, and electron-hole pairs under applied bias
Detectability of microcalcifications and small lesions in mammography has driven the development of high spatial
resolution imagers with small pixel pitch. In this work, we study the detector resolution limits of amorphous selenium (a-
Se) with a detailed Monte Carlo transport code for simulation of direct x-ray detectors. The model takes into account
generation and re-absorption of characteristic x rays, spreading due to Compton scattering and high-energy secondary
electron transport, and drift and diffusion of electron-hole pairs under the applied external electric field. The transport of
electron-hole pairs is achieved with a spatiotemporal model that accounts for recombination and trapping of carriers and
Coulombic effects of 3D spatial charge distribution. The location information for each detected electron and hole over
millions of simulation histories are used to build the detector point response. A range of incident x-ray energies are
simulated from 10 to 100 keV. The simulated detector point response can be used to study the spatial resolution
characteristics of detectors at different energies ranges and for calculation of the modulation transfer function and image
quality metrics.
Yuan Fang,Andreu Badal,Aldo Badano, andKarim S. Karim
"Spatial resolution characteristics of a-Se imaging detectors using
Monte Carlo methods with detailed spatiotemporal transport of x-rays,
electrons, and electron-hole pairs under applied bias", Proc. SPIE 8668, Medical Imaging 2013: Physics of Medical Imaging, 86683R (19 March 2013); https://doi.org/10.1117/12.2007436