You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
6 March 2013Monte Carlo simulation of bowtie filter scatter on a wide-cone low-dose CT system
Knowledge of scatter generated by bowtie filter is crucial for providing artifact free images on the wide-cone low-dose
CT scanners. We investigate and determine the scatter level and artifact generated by the widely used bowtie filter in a
wide-cone low-dose CT system. Our approach is to use Monte Carlo simulation to estimate the scatter level generated by
a bowtie filter made of a material with low atomic number. First, major components of CT systems, such as source, prepatient
collimator, flat filter, bowtie filter, body phantom, and an optional post patient collimator (anti-scatter grid), are
built into a 3D model. The scattered photon fluence and the primary transmitted photon fluence are simulated by
MCNP5 - a Monte Carlo simulation toolkit. With the increased interests in the low dose and wide coverage CT
technology, a tube potential of 80 kVp with more than 10 degree of cone angle is selected. The biased sinogram is
created by superimposing scatter signal generated by the bowtie filter onto the primary x-ray beam signal. Finally,
images with artifacts are reconstructed with the biased signal. Methods to reduce bowtie filter scatter are also discussed
and demonstrated.