You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
19 March 2014Toward a dose reduction strategy using model-based reconstruction with limited-angle tomosynthesis
Model-based iterative reconstruction (MBIR) is an emerging technique for several imaging modalities and appli-
cations including medical CT, security CT, PET, and microscopy. Its success derives from an ability to preserve
image resolution and perceived diagnostic quality under impressively reduced signal level. MBIR typically uses a
cost optimization framework that models system geometry, photon statistics, and prior knowledge of the recon-
structed volume. The challenge of tomosynthetic geometries is that the inverse problem becomes more ill-posed
due to the limited angles, meaning the volumetric image solution is not uniquely determined by the incom-
pletely sampled projection data. Furthermore, low signal level conditions introduce additional challenges due to
noise. A fundamental strength of MBIR for limited-views and limited-angle is that it provides a framework for
constraining the solution consistent with prior knowledge of expected image characteristics. In this study, we
analyze through simulation the capability of MBIR with respect to prior modeling components for limited-views,
limited-angle digital breast tomosynthesis (DBT) under low dose conditions. A comparison to ground truth
phantoms shows that MBIR with regularization achieves a higher level of fidelity and lower level of blurring
and streaking artifacts compared to other state of the art iterative reconstructions, especially for high contrast
objects. The benefit of contrast preservation along with less artifacts may lead to detectability improvement of
microcalcification for more accurate cancer diagnosis.
The alert did not successfully save. Please try again later.
Eri Haneda, J. Eric Tkaczyk, Giovanni Palma, Răzvan Iordache, Scott Zelakiewicz, Serge Muller, Bruno De Man, "Toward a dose reduction strategy using model-based reconstruction with limited-angle tomosynthesis," Proc. SPIE 9033, Medical Imaging 2014: Physics of Medical Imaging, 90330V (19 March 2014); https://doi.org/10.1117/12.2042897