23 July 2014 Reducing noise from a Stirling micro cooler used with an InSb diode
Nicolas R. Bingham, Michael C. B. Ashley
Author Affiliations +
Abstract
Stirling micro coolers, such as the K508 from Ricor, are useful components of scientific instruments when there is a need to remove modest amounts of heat (~1/2W) at liquid nitrogren temperatures with an input power of less than 10W. The action of the cooler can, however, couple noise into sensitive detectors through a variety of mechanisms such as electromagnetic interference, mechanical vibration, and small temperature fluctuations. We report on successful noise-mitigation strategies for our application, an InSb diode for detecting light at 2.4 microns. The largest benefit was obtained by sychronizing the integration times with the position of the piston in the micro cooler. The piston position was determined using a hall-effect rotor position sensor in the driving motor.
© (2014) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Nicolas R. Bingham and Michael C. B. Ashley "Reducing noise from a Stirling micro cooler used with an InSb diode", Proc. SPIE 9154, High Energy, Optical, and Infrared Detectors for Astronomy VI, 91541V (23 July 2014); https://doi.org/10.1117/12.2055551