Paper
2 June 2014 Field test of microbend fiber sensor for hospital use
Author Affiliations +
Proceedings Volume 9157, 23rd International Conference on Optical Fibre Sensors; 91579A (2014) https://doi.org/10.1117/12.2059417
Event: OFS2014 23rd International Conference on Optical Fiber Sensors, 2014, Santander, Spain
Abstract
In this paper, we report a field test of a microbend fiber sensor for simultaneous measurement of breathing rate, breathing pattern, Ballistocardiogram and heart rate during magnetic resonance imaging (MRI). Comparative experiments conducted between our sensor and commercial physiologic device on a healthy male subject showed an accuracy of +/-2bpm for simultaneous measurement of both breathing rate and heart rate. Our preliminary field test on simultaneous measurement of breathing rate and heart rate in a clinical trial conducted on 11 healthy subjects in the 3.0 Tesla MRI environment showed very good agreement compared with measurements obtained from conventional MRcompatible devices.
© (2014) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Zhihao Chen, Doreen Lau, Ju Teng Teo, Soon Huat Ng, Xiufeng Yang, and Pin Lin Kei "Field test of microbend fiber sensor for hospital use", Proc. SPIE 9157, 23rd International Conference on Optical Fibre Sensors, 91579A (2 June 2014); https://doi.org/10.1117/12.2059417
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Heart

Fiber optics sensors

Magnetic resonance imaging

Beam propagation method

Transceivers

Multimode fibers

Back to Top