Paper
17 March 2015 Influence of the grayscale on phantom-based image quality assessment in x-ray computed tomography
F. Noo, K. Hahn, H. C. Davidson, U. Rassner, M. E. Heilbrun
Author Affiliations +
Abstract
Radiation dose associated with CT scans has become an important concern in medical imaging. Fortunately, there are many pathways to reducing dose. A complicated aspect, however, is to ensure that image quality is not affected while reducing the dose. A preferred method to assess image quality is ROC analysis, possibly with a search process. For early assessment of new imaging solutions, utilization of human observers and real patient data is rarely practical. Instead, studies involving phantoms and model observers are often preferred. We present here an experimental result that sheds light on how the grayscale window affects human observer performance in a typical phantom-based study; and we also present an analysis that clarifies how the grayscale window affects the statistics of the image. These studies provide a better understanding of possible consequences associated with not including a grayscale window in studies with model observers, as is typical.
© (2015) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
F. Noo, K. Hahn, H. C. Davidson, U. Rassner, and M. E. Heilbrun "Influence of the grayscale on phantom-based image quality assessment in x-ray computed tomography", Proc. SPIE 9416, Medical Imaging 2015: Image Perception, Observer Performance, and Technology Assessment, 94160G (17 March 2015); https://doi.org/10.1117/12.2082914
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Image quality

X-ray computed tomography

Statistical analysis

Image processing

Computed tomography

Data modeling

Medical imaging

Back to Top